Product Attributes
Gaskets and O-Rings
FDA approved Ethylene Propylene, PTFE encapsulated, Silicone or Nitrile.
Maximum Differential Pressure
Normal flow direction at: |
20°C (68°F): |
6.0bar (87psi) |
80°C (176°F): |
4.0bar (58psi) |
100°C (212°F): |
3.0bar (43psi) |
120°C (248°F): |
2.0bar (29psi) |
Integrity Testing
Each Biofil™ Plus module of every cartridge is individually integrity tested using the Diffusive Flow Test, which correlates to the HIMA and ASTM F838-05 bacterial challenge tests. Non-destructive integrity tests, such as Pressure Hold, Diffusive Flow and Bubble Point, can be performed by customers. Procedural details are available from Porvair.
Effective Filtration Area
Pore Size Rating |
Effective Filtration Area (each 254mm (10″) module) |
0.2, 0.45 and 0.65μm |
0.48m2 (5.2ft2) |
Operating Temperature
Maximum continuous: 60°C (140°F)
Applications
- Biopharmaceuticals
For the sterilisation of biological fluids, cell culture media, sera and blood fractionations.
- Fermentation
For providing sterile feed stock for the production of antibiotics and enzymes.
- Opthalmic solutions
- API’s
For the clarification and sterilisation of a wide range of active pharmaceutical ingredients.
- LVP’s
For final filtration of Total Nutritional Fluids, dextrose, amino acids and saline solutions.
- Beverages
Clean Water Flow Rates